GSP-100 Sampling Probe

Description

The unique sampling probe design is applied to a variety of working conditions (high temperature, high humidity, high dust, high corrosion, etc.), supported with a variety of sampling tube, filter, heating units, etc..It has the feature of large filter area and capacity, good air tightness, high filtration precision, leak-proof design, low gas resistance; high precision temperature control, temperature overload protection, reliable operation, fast response, easy installation and maintenance.

Spesification

Sampling Temperature≤600℃
Sampling Chamber Heating Temp.120±10℃
Filtration Precision2μm
Temperature Accuracy±0.5℃
Working Temperature-10℃~+60℃
Dielectric StrengthGB 6587.7
Blowback gas source(0.4~0.8)MPa N2, clean instrument air

Features

  • Titanium alloy filter provides high corrosion resistance in high humidity, high temperature environment
  • Unique internal structure design reduces effectively the dead space of the sample gas
  • Filter uses a sintered alloy to filter 2μ m and more dust particles and small gas resistance
  • Efficient inside and outside blowback structure design can effectively remove the dust, avoid congestion, reduce daily maintenance

Application

Apply to continuous acquisition of exhaust gas samples after the combustion process, typically high dust and moisture content of the sample gas collection in power plant flue gas, cement production process