Description
AQMS-600 nitrogen oxides (NO-NO2–NOx) analyzer utilizes chemiluminescence technology indicated by US EPA as the federal reference method for monitoring multiple forms of nitrogen oxides.
AQMS-600 nitrogen oxides (NO-NO2–NOx) analyzer utilizes chemiluminescence technology indicated by US EPA as the federal reference method for monitoring multiple forms of nitrogen oxides.
Principle
|
Chemiluminiscence
|
Standard Range
|
Max:0~20ppm
Min:0~100ppb (Selectable)
|
Zero Noise
|
≤0.2ppb ( RMS )
|
Span Noise
|
≤2.5ppb ( RMS )
|
Lower Detectable Limit
|
≤0.4ppb ( RMS )
|
Zero Drift
|
1ppb/24h
|
Span Drift
|
±5ppb/24h
|
Linearity
|
<1%F.S. |
Precision
|
<1%
|
Response Time
|
T90<60s
|
Power Requirement | 220V AC/50Hz; 110V AC/60Hz |