AQMS-500 sulfur dioxide (SO2) analyzer applies UV fluorescence technology with the photomultiplier tube (PMT) to measure SO2.
AQMS-500 sulfur dioxide (SO2) analyzer applies UV fluorescence technology with the photomultiplier tube (PMT) to measure SO2.
Principle | UV Fluorescence |
Standard Range | Max:0~20ppm Min:0~100ppb (Selectable) |
Zero Noise | ≤0.25ppb (RMS) |
Span Noise | ≤2.5ppb (RMS) |
Display | Digital |
Lower Detectable Limit | 0.5ppb |
Zero Drift | <1ppb/24h |
Span Drift | ±5ppb/24h |
Linearity | <1%F.S. |
Precision | <1% |
Response Time | T90<120s |
Sample Flow Rate | (650±65)sccm |
Data Transmission | 2 channel analog (4~20) mA; 12 digital input/output; 4-way relay output; |
Calibration | Multi-point calibrator |
Communication | RS232/RS485/Ethernet |
Operating Temperature | US EPA Specification 20~30°C, Actual applicable:0~40°C |
Operating Humidity | 0~95%RH(No:condensation) |
Power Requirement | 220V AC/50Hz; 110V AC/60Hz |